Multivariate calibration for the improvement of the quantification of cadmium in the presence of potassium as interferent by total reflection X-ray fluorescence

Medina González G.; Castillo R.D.P.; Neira J.Y.

Abstract

A multivariate calibration method was applied to improve the accuracy of the determination of cadmium in the presence of potassium by total reflection X-ray fluorescence (TXRF). Due to TXRF's relatively low resolution, a particular interference caused by potassium at line K alpha 3.31 keV as interference can compromise the cadmium determination at line L alpha 3.13 keV, respectively. The method is based on the hypothesis that the application of a multivariate calibration method, for example, partial least squares, could reduce variations due to interference and, consequently, improve the selectivity and accuracy in Cd determination. In this work, this strategy was evaluated on the X-ray fluorescence emission signals between 2.50 and 3.90 keV for a set of 26 different synthetic calibration mixtures and eight different mixtures for external calibration. Based on a certified reference material (Clay 2-CRM051) with high potassium levels, a significant improvement of accuracy for cadmium determination was observed, overcoming the problems associated with spectral interferences by potassium.

Más información

Título según WOS: Multivariate calibration for the improvement of the quantification of cadmium in the presence of potassium as interferent by total reflection X-ray fluorescence
Título según SCOPUS: Multivariate calibration for the improvement of the quantification of cadmium in the presence of potassium as interferent by total reflection X-ray fluorescence
Título de la Revista: X-RAY SPECTROMETRY
Volumen: 48
Número: 6
Editorial: Wiley
Fecha de publicación: 2019
Página de inicio: 700
Página final: 707
Idioma: English
DOI:

10.1002/xrs.3113

Notas: ISI, SCOPUS