Radiation dose and image quality for paediatric interventional cardiology

Vañó E.; Ubeda, C.; Leyton, F.; Miranda, P.

Abstract

Radiation dose and image quality for paediatric protocols in a biplane x-ray system used for interventional cardiology have been evaluated. Entrance surface air kerma (ESAK) and image quality using a test object and polymethyl methacrylate (PMMA) phantoms have been measured for the typical paediatric patient thicknesses (4-20 cm of PMMA). Images from fluoroscopy (low, medium and high) and cine modes have been archived in digital imaging and communications in medicine (DICOM) format. Signal-to-noise ratio (SNR), figure of merit (FOM), contrast (CO), contrast-to-noise ratio (CNR) and high contrast spatial resolution (HCSR) have been computed from the images. Data on dose transferred to the DICOM header have been used to test the values of the dosimetric display at the interventional reference point. ESAK for fluoroscopy modes ranges from 0.15 to 36.60 μGy/frame when moving from 4 to 20 cm PMMA. For cine, these values range from 2.80 to 161.10 μGy/frame. SNR, FOM, CO, CNR and HCSR are improved for high fluoroscopy and cine modes and maintained roughly constant for the different thicknesses. Cumulative dose at the interventional reference point resulted 25-45% higher than the skin dose for the vertical C-arm (depending of the phantom thickness). ESAK and numerical image quality parameters allow the verification of the proper setting of the x-ray system. Knowing the increases in dose per frame when increasing phantom thicknesses together with the image quality parameters will help cardiologists in the good management of patient dose and allow them to select the best imaging acquisition mode during clinical procedures. © 2008 Institute of Physics and Engineering in Medicine.

Más información

Título según WOS: Radiation dose and image quality for paediatric interventional cardiology
Título según SCOPUS: Radiation dose and image quality for paediatric interventional cardiology
Título de la Revista: PHYSICS IN MEDICINE AND BIOLOGY
Volumen: 53
Número: 15
Editorial: IOP PUBLISHING LTD
Fecha de publicación: 2008
Página de inicio: 4049
Página final: 4062
Idioma: English
URL: http://stacks.iop.org/0031-9155/53/i=15/a=003?key=crossref.ac5881379a0bb936ba59f158ca681cbb
DOI:

10.1088/0031-9155/53/15/003

Notas: ISI, SCOPUS