Field-based cavity ring-down spectrometry of 13C in soil-respired CO2
Abstract
Measurement of soil-respired CO2 at high temporal resolution and sample density is necessary to accurately identify sources and quantify effluxes of soil-respired CO2. A portable sampling device for the analysis of C-13 values in the field is described herein. CO2 accumulated in a soil chamber was batch sampled sequentially in four gas bags and analysed by Wavelength-Scanned Cavity Ring-down Spectrometry (WS-CRDS). A Keeling plot (1/[CO2] versus C-13) was used to derive C-13 values of soil-respired CO2. Calibration to the C-13 Vienna Peedee Belemnite scale was by analysis of cylinder CO2 and CO2 derived from dissolved carbonate standards. The performance of gas-bag analysis was compared to continuous analysis where the WS-CRDS analyser was connected directly to the soil chamber. Although there are inherent difficulties in obtaining absolute accuracy data for C-13 values in soil-respired CO2, the similarity of C-13 values obtained for the same test soil with different analytical configurations indicated that an acceptable accuracy of the C-13 data were obtained by the WS-CRDS techniques presented here. Field testing of a variety of tropical soil/vegetation types, using the batch sampling technique yielded C-13 values for soil-respired CO2 related to the dominance of either C-3 (tree, C-13=-27.8 to-31.9 parts per thousand) or C-4 (tropical grass, C-13=-9.8 to-13.6 parts per thousand) photosynthetic pathways in vegetation at the sampling sites. Standard errors of the Keeling plot intercept C-13 values of soil-respired CO2 were typically<0.4 parts per thousand for analysis of soils with high CO2 efflux (>7-9molm(-2)s(-1)).
Más información
Título según WOS: | ID WOS:000320017800007 Not found in local WOS DB |
Título de la Revista: | ISOTOPES IN ENVIRONMENTAL AND HEALTH STUDIES |
Volumen: | 49 |
Número: | 2 |
Editorial: | TAYLOR & FRANCIS LTD |
Fecha de publicación: | 2013 |
Página de inicio: | 232 |
Página final: | 242 |
DOI: |
10.1080/10256016.2013.750606 |
Notas: | ISI |