Serotonin secretion by blood platelets: accuracy of high-performance liquid chromatography-electrochemical technique compared with the isotopic test and use in a clinical laboratory
Abstract
--- - "Background: Mild secretion defects are the most frequent and challenging blood platelet disorders to diagnose. Most d-granule secretion tests lack validation, are not quantitative, or have unreliable response to weak platelet agonists." - "Objectives: To compare platelet serotonin secretion by HPLC-electrochemical detection technique (HPLC-ECD) with the reference isotopic test (3H-5-HT), evaluating its performance in clinical laboratories." - "Methods: The assay validation followed STARD-2015 recommendations. HPLC-ECD measured the nonsecreted serotonin remaining in platelet pellets after aggregation, comparing it with the reference H-3-5-HT assay. We studied subjects with inherited and aspirin-induced blood platelet disorders and assessed the HPLC-ECD operation for routine clinical diagnosis." - "Results: Calibration curves were linear (R-2 = 0.997), with SD for residuals of 3.91% and analytical sensitivity of 5ng/mL. Intra- and interassay imprecision bias ranged between -8.5% and 2.1% and -9% and 3.1%, respectively. Serotonin recovery and stability were >95%, and the variability range of measurements was -5.5% to 4.6%. Statistical differences detected between tests were biologically irrelevant, with bias of 1.48% (SD, 8.43) and CI agreement of -18% to 15%. Both assays distinctly detected platelet secretion induced by 10 mu M epinephrine and 4 mu mM adenosine diphosphate. However, HPLC-ECD is quantitative and more sensitive to low serotonin content in blood platelets. Reference cutoffs for each agonist were determined in 87 subjects. Initially, the HPLC-ECD requires relatively expensive equipment and trained operators but has remarkably cheap running costs and a turn-around time of 24-36 hours. We have used this diagnostic tool routinely for >8 years." - "Conclusion: HPLC-ECD assay for platelet serotonin secretion is highly accurate, has advantages over the reference H-3-5-HT test, and is suitable as a clinical laboratory technique."
Más información
Título según WOS: | Serotonin secretion by blood platelets: accuracy of high-performance liquid chromatography-electrochemical technique compared with the isotopic test and use in a clinical laboratory |
Título de la Revista: | RESEARCH AND PRACTICE IN THROMBOSIS AND HAEMOSTASIS |
Volumen: | 7 |
Número: | 5 |
Editorial: | Elsevier |
Fecha de publicación: | 2023 |
DOI: |
10.1016/j.rpth.2023.102156 |
Notas: | ISI |