Determination of Cd and Pb at ?g/L levels by HHPN-beam injection flame furnace-AAS

Neira, J; Berndt, H

Keywords: article

Abstract

Flame-heated tubes are widespread in flame-AAS, mainly for the determination of hydride-forming elements. Instead of the introduction of gaseous compounds liquids can also be introduced continuously in such an absorption cell. With the aid of an HPLC pump the liquid is forced through a very fine nozzle, generating an aerosol beam less than 0.5mm in diameter. This beam travels a distance of 10cm as a "free-flying aerosol jet" into the sample introduction hole of a metal tube furnace placed in the flame. Both introduction of the entire sample and the long residence time lead to a considerable improvement in power of detection. The detection limit for 100 ?L samples amounts to 7 ?g/L (Pb) and to 0. 2 ?g/L (Cd), which means an increase in power of detection of between one and two orders of magnitude compared to conventional flame-AAS. The relative standard deviation (100 ?L sample volume, N = 10) was found to be 1.3% (signal area) for 600 ?g/L Pb and to be 1.5% (signal-area) for 15 ?g/L Cd. RSD values from measurements in peak-height amounted to 2.2% (Pb) and to 1.7% (Cd). © Springer-Verlag 2000.

Más información

Título según SCOPUS: Determination of Cd and Pb at ?g/L levels by HHPN-beam injection flame furnace-AAS
Título de la Revista: Fresenius' Journal of Analytical Chemistry
Volumen: 368
Número: 7
Editorial: Springer-Verlag
Fecha de publicación: 2000
Página de inicio: 649
Página final: 655
Idioma: English
URL: http://www.scopus.com/inward/record.url?eid=2-s2.0-2042430949&partnerID=q2rCbXpz