Absolute Calibration of Proportional Counter Based Fast Pulsed Neutron Detectors with Resolution Below 10(5) neutron/pulse

Tarifeno-Saldivia, A; Mayer, RE; Pavez, C.; Soto L.

Abstract

A method for absolute calibration of proportional counters for pulsed fast neutrons is presented. The method is based on the use of an isotopic standard source and development of a model for counting detected events from area of a signal compounded by single piled up neutron pulses. Effects of detection counting statistics and electrical background noise are also considered. The method is applied in detectors used for D-D neutron yield measurements in low emission plasma focus devices.

Más información

Título según WOS: Absolute Calibration of Proportional Counter Based Fast Pulsed Neutron Detectors with Resolution Below 10(5) neutron/pulse
Título según SCOPUS: Absolute calibration of proportional counter based fast pulsed neutron detectors with resolution below 105 neutron/pulse
Título de la Revista: VIII INTERNATIONAL CONGRESS OF ENGINEERING PHYSICS
Volumen: 511
Número: 1
Editorial: IOP PUBLISHING LTD
Fecha de publicación: 2014
Idioma: English
DOI:

10.1088/1742-6596/511/1/012087

Notas: ISI, SCOPUS