In silicon evaluation of a control system and algorithm for automated insulin infusion in the ICU setting

Ortiz, J.; Guarini, M.; Borzone, G.; Olmos, P.

Abstract

Background It is known that tight control of glucose in the Intensive Care Unit reduces morbidity and mortality not only in diabetic patients but also in those non-diabetics who become transiently hyperglycemic. Taking advantage of a recently marketed subcutaneous glucose sensor we designed an Automatic Insulin Infusion System (AIIS) for inpatient treatment, and tested its stability under simulated clinical conditions. Methods The system included: reference glucose, glucose sensor, insulin and glucose infusion controllers and emergency infusion logic. We carried out computer simulations using Matlab/Simulink®, in both common and worst-case conditions. Results The system was capable of controlling glucose levels without entering in a phase of catastrophic instability, even under severe simulated challenges. Care was taken to include in all simulations the 5-10 minute delay of the subcutaneous glucose signal when compared to the real-time serum glucose signal, a well-known characteristic of all subcutaneous glucose sensors. Conclusions When tested in-Silico, a commercially available subcutaneous glucose sensor allowed the stable functioning of a proportional-derivative Automatic Insulin Infusion System, which was able to maintain glucose within acceptable limits when using a well-established glucose response model simulating a patient. Testing of the system in vivo using animal models is now warranted.

Más información

Título de la Revista: BIOMEDICAL ENGINEERING ONLINE
Volumen: 9
Número: 35
Editorial: BMC
Fecha de publicación: 2010
Idioma: English