Amperometric biosensors for detection of sugars based on the electrical wiring of different pyranose oxidases and pyranose dehydrogenases with osmium redox polymer on graphite electrodes

Tasca, Federico; Timur, Suna; Ludwig, Roland; Haltrich, Dietmar; Volc, Jindrich; Antiochia, Riccarda; Gorton, Lo

Abstract

Electrical wiring of different types of pyranose oxidase (P20) (fungal wild type, recombinant wild type with a hexahistidine tag, mutant form E542K with a hexa-histidine tag) from Trametes multicolor, and recombinant P20 from Coriolus sp. overexpressed in Escherichia coli as well as of pyranose dehydrogenase (PDH) from Agaricus meleagris and Agaricus xanthoderma with an osmium redox polymer (poly(1-vinylimidazole)(12)-[Os(4,4'-dimethyl-2,2'-dipyridyl)(2)Cl2+/+) on graphite electrodes was carried out. After optimization studies using glucose as substrate, the biosensors, which showed the best characteristics in terms of linear range, detection limit and sensitivity were selected, viz. wild type P20 from T multicolor and PDH from A. meleagris. These two enzymes were used and investigated for their selectivity for a number of different sugars.

Más información

Título según WOS: ID WOS:000244220700025 Not found in local WOS DB
Título de la Revista: ELECTROANALYSIS
Volumen: 19
Número: 2-3
Editorial: WILEY-V C H VERLAG GMBH
Fecha de publicación: 2007
Página de inicio: 294
Página final: 302
DOI:

10.1002/elan.200603740

Notas: ISI