Increasing amperometric biosensor sensitivity by length fractionated single-walled carbon nanotubes
Abstract
In this work the sensitivity-increasing effect of single-walled carbon nanotubes (SWCNTs) in amperometric biosensors, depending on their average length distribution, was studied. For this purpose the SWCNTs were oxidatively shortened and subsequently length separated by size exclusion chromatography. Transmission electron micrographs of different fractions of SWCNTs were collected. Diaphorase "wired" to an osmium redox polymer was blended with the shortened SWCNTs of different lengths. Depending on the average length of the SWCNTs the sensitivity of the amperometric biosensor model system towards oxidation of 1.4-dihydronicotinamide adenine dinucleotide (NADH) was increased by a factor of five. The best performance was achieved with SWCNTs of medium length. The linear range for NADH detection was between 5 mu M and 7 mM, the maximum sensitivity was 47 nA mu M-1 cm(-2), and the detection limit was 1 mu M. The biosensor exhibited excellent electrocatalytic properties. Even at relatively high NADH concentrations the oxidative current was limited by the diffusion rate of NADH. (C) 2008 Elsevier B.V. All rights reserved.
Más información
Título según WOS: | ID WOS:000259711900017 Not found in local WOS DB |
Título de la Revista: | BIOSENSORS BIOELECTRONICS |
Volumen: | 24 |
Número: | 2 |
Editorial: | ELSEVIER ADVANCED TECHNOLOGY |
Fecha de publicación: | 2008 |
Página de inicio: | 272 |
Página final: | 278 |
DOI: |
10.1016/j.bios.2008.03.038 |
Notas: | ISI |