Standardless semi-quantitative analysis by PIXE

Rodriguez, T.; Limandri, S.; Suarez, S.; Ortega-Feliu, I.; Trincavelli, J.

Abstract

A method based on the refinement of atomic and experimental parameters developed for the description of PIXE spectra is presented and applied to standardless semi-quantitative PIXE analysis. This method was implemented in the sofware PAMPA (Parameter Assessment Method for PIXE Analysis) and consists in minimizing the quadratic differences between an experimental spectrum and an analytical function proposed to describe it. The first results of PAMPA are presented for the quantification of synthetic and mineral, thin and bulk samples, and they are compared with the results obtained with a commercial software.

Más información

Título según WOS: ID WOS:000401119500013 Not found in local WOS DB
Título de la Revista: JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
Volumen: 32
Número: 5
Editorial: ROYAL SOC CHEMISTRY
Fecha de publicación: 2017
Página de inicio: 1020
Página final: 1030
DOI:

10.1039/c7ja00068e

Notas: ISI