High-power and high-accuracy integrating sphere radiometer: design, characterization, and calibration
Abstract
We describe the design, characterization, and calibration of a high-power and high-accuracy transfer standard for optical power measurements in fibers based on an integrating sphere radiometer working from -50 to +30 dBm. The integrating sphere radiometer has been calibrated in the spectral range 1250-1650 nm by use of an electrically calibrated pyroelectric radiometer and four tunable laser diodes. The total uncertainty obtained is less than +/- 0.8% for these wavelength and power ranges. (c) 2006 Optical Society of America.
Más información
Título según WOS: | ID WOS:000234878100015 Not found in local WOS DB |
Título de la Revista: | APPLIED OPTICS |
Volumen: | 45 |
Número: | 3 |
Editorial: | OPTICAL SOC AMER |
Fecha de publicación: | 2006 |
Página de inicio: | 511 |
Página final: | 518 |
DOI: |
10.1364/AO.45.000511 |
Notas: | ISI |