High-power and high-accuracy integrating sphere radiometer: design, characterization, and calibration

Carrasco-Sanz, A; Martin-Lopez, S; Corredera, P; Gonzalez-Herraez, M; Hernanz, ML

Abstract

We describe the design, characterization, and calibration of a high-power and high-accuracy transfer standard for optical power measurements in fibers based on an integrating sphere radiometer working from -50 to +30 dBm. The integrating sphere radiometer has been calibrated in the spectral range 1250-1650 nm by use of an electrically calibrated pyroelectric radiometer and four tunable laser diodes. The total uncertainty obtained is less than +/- 0.8% for these wavelength and power ranges. (c) 2006 Optical Society of America.

Más información

Título según WOS: ID WOS:000234878100015 Not found in local WOS DB
Título de la Revista: APPLIED OPTICS
Volumen: 45
Número: 3
Editorial: OPTICAL SOC AMER
Fecha de publicación: 2006
Página de inicio: 511
Página final: 518
DOI:

10.1364/AO.45.000511

Notas: ISI