Study of out-of-field dose in photon radiotherapy: a commercial treatment planning system versus measurements and Monte Carlo simulations.

Sánchez-Nieto, Beatriz; Medina-Ascanio, Karem N.; Rodríguez-Mongua, José L.; Doerner, Edgardo; Espinoza, Ignacio

Abstract

Purpose: An accurate assessment of out-of-field dose is necessary to estimate the risk of second cancer after radiotherapy and the damage to the organs at risk surrounding the planning target volume. Although treatment planning systems (TPSs) calculate dose distributions outside the treatment field, little is known about the accuracy of these calculations. The aim of this work is to thoroughly compare the out-of-field dose distributions given by two algorithms implemented in the Monaco TPS, with measurements and full Monte Carlo simulations. Methods: Out-of-field dose distributions predicted by the Collapsed Cone Convolution (CCC) and Monte Carlo (MCMonaco) algorithms, built into the commercially available Monaco version 5.11 TPS, are compared with measurements carried out on an Elekta Axesse linear accelerator. For the measurements, ion chambers, thermoluminescent dosimeters, and EBT3 film are used. The BEAMnrc code, built on the EGSnrc system, is used to create a model of the Elekta Axesse with the Agility collimation system, and the space phase file generated is scored by DOSXYZnrc to generate the dose distributions (MCEGSnrc). Three different irradiation scenarios are considered: i) a 10x10 cm2 field, ii) an IMRT prostate plan, and iii) a 3-field lung plan. Monaco’s calculations, experimental measurements, and Monte Carlo simulations are carried out in water and/or in an ICRP110 phantom. Results For the 10x10 cm2 field case, CCC underestimated the dose, compared to ion chamber measurements, by 13% (differences relative to the algorithm) on average between the 5% and the 2% isodoses. MCMonaco underestimated the dose only from approximately the 2% isodose for this case. Qualitatively similar results were observed for the studied IMRT case when compared to film dosimetry. For the 3-field lung plan, dose underestimations of up to 90% for MCMonaco and 60% for CCC, relative to MCEGSnrc simulations, were observed in mean dose to organs located beyond the 2% isodose. Conclusions: This work shows that Monaco underestimates out-of-field doses in almost all the cases considered. Thus, it does not describe dose distribution beyond the border of the field accurately. This is in agreement with previously published works reporting similar results for other TPSs. Analytical models for out-of-field dose assessment, MC simulations or experimental measurements may be an adequate alternative for this purpose.

Más información

Título de la Revista: MEDICAL PHYSICS
Editorial: Wiley
Fecha de publicación: 2020
DOI:

MP14356

Notas: ISI SCOPUS