Reliability of Low-Cost Near-Infrared Spectroscopy in the Determination of Muscular Oxygen Saturation and Hemoglobin Concentration during Rest, Isometric and Dynamic Strength Activity

Miranda-Fuentes, Claudia; Maria Guisado-Requena, Isabel; Delgado-Floody, Pedro; Arias-Poblete, Leonidas; Perez-Castilla, Alejandro; Jerez-Mayorga, Daniel; Javier Chirosa-Rios, Luis

Abstract

Background: The objective of this study was to establish the reliability of the Humon Hex near-infrared reflectance spectroscopy (NIRS) in determining muscle oxygen saturation (SmO2) and hemoglobin concentration (Hgb) at rest and during isometric and dynamic strength exercises using a functional electromechanical dynamometer (FEMD). Methods: The SmO2 and Hgb values of sixteen healthy adults (mean +/- standard deviation (SD): Age = 36.1 +/- 6.4 years) were recorded at rest and during isometry (8 s), dynamic strength I (initial load of 40% of the average isometric load, with 2 kg increments until muscle failure) and dynamic strength II (same as I, but with an initial load of 40% of the maximum isometric load) activity. To evaluate the reliability in the determination of SmO2 and Hgb of this device, intraclass correlation coefficient (ICC), standard error of measurement (SEM) and coefficient of variation (CV) were obtained. Results: The main results obtained are SmO2 at rest (CV = 5.76%, SEM = 3.81, ICC = 0.90), isometric strength (CV = 3.03%, SEM = 2.08, ICC = 0.92), dynamic strength I (CV = 10.6, SEM = 7.17, ICC = 0.22) and dynamic strength II (CV = 9.69, SEM = 6.75, ICC = 0.32); Hgb at rest (CV = 1.97%, SEM = 0.24, ICC = 0.65), isometric strength (CV = 0.98%, SEM = 0.12, ICC = 0.96), dynamic strength I (CV = 3.25, SEM = 0.40, ICC = 0.54) and dynamic strength II (CV = 2.74, SEM = 0.34, ICC = 0.65). Conclusions: The study shows that Humon Hex is a reliable device to obtain SmO2 and Hgb data in healthy adult subjects at rest and during isometric strength, providing precision for measurements made with this device.

Más información

Título según WOS: Reliability of Low-Cost Near-Infrared Spectroscopy in the Determination of Muscular Oxygen Saturation and Hemoglobin Concentration during Rest, Isometric and Dynamic Strength Activity
Título de la Revista: International #Journal of Environmental Research and Public Health
Volumen: 17
Número: 23
Editorial: MDPI
Fecha de publicación: 2020
DOI:

10.3390/IJERPH17238824

Notas: ISI