The effect and detection of organic contamination on the measured hydrogen and oxygen stable isotopic composition of water vapour by laser spectroscopy methods

Millar, Cody; Janzen, Kim; Nehemy, Magali; Koehler, Geoff; Hervé-Fernández, Pedro; McDonnell, JJ

Keywords: contamination, stable isotopes, Oxyge-17, Flagging

Abstract

RATIONALE Hydrogen and oxygen stable isotope ratios (δ2H, δ17O, and δ18O values) are commonly used tracers of water. These ratios can be measured by isotope ratio infrared spectroscopy (IRIS). However, IRIS approaches are prone to errors induced by organic compounds present in plant, soil, and natural water samples. A novel approach using 17O-excess values has shown promise for flagging spectrally-contaminated plant samples during IRIS analysis. A systematic assessment of this flagging system is needed to prove it useful. METHODS Errors induced by methanol and ethanol water mixtures on measured IRIS and IRMS results were evaluated. For IRIS analyses both liquid and vapour-mode (via direct vapour equilibration) methods are used. The δ2H, δ17O, and δ18O values were measured and compared with known reference values to determine the errors induced by methanol and ethanol contamination. In addition, the 17O-excess contamination detection approach was tested. This is a post processing detection tool for both liquid and vapour IRIS triple-isotope analyses, utilizing calculated 17O-excess values to flag contaminated samples. RESULTS Organic contamination induced significant errors in IRIS results which were not seen in IRMS results. Methanol caused larger errors than ethanol. Results from vapour-IRIS analyses had larger errors than those from liquid-IRIS analyses. The 17O-excess approach identified methanol driven error in liquid and vapour-mode IRIS samples at levels where isotope results became unacceptably erroneous. For ethanol contaminated samples a mix of erroneous and correct flagging occurred with the 17O-excess method. Our results indicate that methanol is the more problematic contaminant for data corruption. The 17O-excess method was therefore useful for data quality control. CONCLUSION Organic contamination caused significant errors in IRIS stable isotope results. These errors were larger during vapour analysis than during liquid IRIS analyses, and larger for methanol than ethanol contamination. The 17O-excess method is highly sensitive for detecting narrowband (methanol) contamination error in vapour and liquid analysis modes on IRIS.

Más información

Título de la Revista: RAPID COMMUNICATIONS IN MASS SPECTROMETRY
Editorial: Wiley
Fecha de publicación: 2021
Idioma: Inglés
URL: https://analyticalsciencejournals.onlinelibrary.wiley.com/doi/10.1002/rcm.9118