Effects of osseodensification protocol on insertion, removal torques, and resonance frequency analysis of BioHorizons® conical implants. An ex vivo study
Abstract
Objective: Quantify the effect of the osseodensification Densah® protocol on the insertion torque, ISQ, and the removal torque of conical BioHorizons® implants. Materials and methods: An ex vivo model over fresh pig tibia bone was used. Test group (TG) included 50 osteotomies using Densah® osseodensification protocol, and the control group (CG), 50 osteotomies using BioHorizons®âs recommended procedure. Conical BioHorizons® implants (3.8 à 10.5 mm) were implanted, verifying the insertion torque with a manual torque meter. ISQ values were registered with Ostell® device. Finally, implants were removed with manual reverse torque registering the values. Results were analyzed and compared with the Mann-Whitney test and t-test. Results: Median and interquartile range per group were as follows: insertion torque, CG: 26 (12) Ncm; TG: 42 (26) Ncm, removal torque, CG: 25 (20) Ncm; TG: 40 (28) Ncm, ISQ value, CG: 69.25 (5.5); TG: 71.5 (4). All variables were significantly higher (p ⤠0.05) in the osseodensification group. Conclusions: The Osseodensification technique may improve primary stability in the clinical scenario on tapered implants. Further human RCTs are necessary to validate this.
Más información
| Título según SCOPUS: | Effects of osseodensification protocol on insertion, removal torques, and resonance frequency analysis of BioHorizons® conical implants. An ex vivo study |
| Título de la Revista: | Journal of Oral Biology and Craniofacial Research |
| Volumen: | 10 |
| Número: | 4 |
| Editorial: | Elsevier B.V. |
| Fecha de publicación: | 2020 |
| Página final: | 628 |
| Idioma: | English |
| DOI: |
10.1016/j.jobcr.2020.08.019 |
| Notas: | SCOPUS |