Comparison of Improved Unidirectional Dual Velocity-Encoding MRI Methods
Abstract
Background: In phase-contrast (PC) MRI, several dual velocity encoding methods have been proposed to robustly increase velocity-to-noise ratio (VNR), including a standard dual-VENC (SDV), an optimal dual-VENC (ODV), and bi- and triconditional methods. Purpose: To develop a correction method for the ODV approach and to perform a comparison between methods. Study Type: Casecontrol study. Population: Twenty-six volunteers. Field Strength/Sequence: 1.5 T phase-contrast MRI with VENCs of 50, 75, and 150 cm/second. Assessment: Since we acquired single-VENC protocols, we used the background phase from high-VENC (VENC
Más información
| Título según WOS: | Comparison of Improved Unidirectional Dual Velocity-Encoding MRI Methods |
| Título según SCOPUS: | Comparison of Improved Unidirectional Dual Velocity-Encoding MRI Methods |
| Título de la Revista: | Journal of Magnetic Resonance Imaging |
| Volumen: | 57 |
| Número: | 3 |
| Editorial: | John Wiley and Sons Inc. |
| Fecha de publicación: | 2023 |
| Página de inicio: | 763 |
| Página final: | 773 |
| Idioma: | English |
| DOI: |
10.1002/jmri.28305 |
| Notas: | ISI, SCOPUS |