Photoplethysmogram Fits Finger Blood Pressure Waveform for non-Invasive and minimally-Intrusive Technologies Evaluation of Derivative Approaches
Keywords: VPG, nImI, Arterial Hypertension, Fractional Derivatives, Finapres, Photoplethysmography Derivatives.
Abstract
The purpose of this work is to fit Photoplethysmography (PPG) to finger Arterial Pressure (fiAP) waveform using derivative approaches. Derivative approaches consider using Linear Combination of Derivatives (LCD) and Fractional Derivatives (FDPα). Four informed healthy subjects, aging 35.8 ± 11.0 years old, agreed to perform Handgrip maneuvers. Signals are recorded continually; a Finapres NOVA device is used for fiAP, while a BIOPAC System is used for PPG and ECG. PPG is smoothed and segmented by heartbeat; recording sections interfered with spiky blocking noise, are eliminated. Finally, PPG is processed using LCD and FDPα and their results are enriched using Lasso technique. Twenty records per subject at rest and twenty at raised BP are analyzed. Results show PPG to fiAP fitting errors 5.38% ± 0.91 at resting fiAP and 5.86% ± 1.21 at raised fiAP, being always lower than 15%, suggesting that derivative approaches could be suitable for medical applications.
Más información
Título de la Revista: | BIOSIGNALS |
Número: | 17 |
Editorial: | SCITEPRESS |
Fecha de publicación: | 2017 |
Página de inicio: | 155 |
Página final: | 162 |
Idioma: | Español |
DOI: |
Proceedings of the 10th International Joint Conference on Biomedical Engineering Systems and Technologies |