Validity and reliability of the haefni health system 1.0 device in the measurement of the isokinetic velocity range Validación y fiabilidad del dispositivo haefni health system 1.0 en la medición de la velocidad en el rango isocinético
Abstract
When rating any assessment protocol one of the key issues is to determine the validity and reliability of the devices used. The aim of this study was to analyze the validity and reliability of the variable velocity isokinetic range (VRI) in Haefni Health System (HHe 1.0). Two assessment protocols were design to test bot, validity and reliability. The results showed high Index of Intraclass Correlation (ICC2,1) for measures of validity and reliability of the VRI (0,998 and 0,99, respectively). Moreover, when the reliability was assessed at each criterion velocity high values of ICC2,1, range 0,71-0,99 and low CV and SEM, range 0,19-3,73% and range 0.001- 0.004 m.s-1, were found respectively. The results obtained confirm the validity and reliability of HHe1.0 device for measuring VRI variable. © 2014: Servicio de Publicaciones de la Universidad de Murcia (España).
Más información
Título según SCOPUS: | Validity and reliability of the haefni health system 1.0 device in the measurement of the isokinetic velocity range [Validación y fiabilidad del dispositivo haefni health system 1.0 en la medición de la velocidad en el rango isocinético] |
Título de la Revista: | Cuadernos de Psicologia del Deporte |
Volumen: | 14 |
Fecha de publicación: | 2014 |
Página de inicio: | 91 |
Página final: | 98 |
DOI: |
10.4321/S1578-84232014000200010 |
Notas: | SCOPUS |