Instrumental planar chromatographic method for determination of carbamazepine in human serum
Abstract
An instrumental planar chromatographic (HPTLC) method for quantification of carbamazepine in human serum was developed using liquid-liquid extraction with dichloromethane, fluorescence activation with perchloric acid 60%/ethanol/water (1:1:1, v/v) and fluorescence detection. Planar chromatographic separation was performed on precoated silica gel F254 HPTLC plates using a mixture of ethyl acetate/toluene/methanol/acetic acid glacial (5:4:0.5:0.5, v/v) as mobile phase. Densitometric detection was done at 366 nm. The method was validated for linearity, precision and accuracy. Linear calibration curves in the range of 3 and 20 ng/µL showed correlation coefficient of 0.998. The intra-assay and inter-assay precision, expressed as the RSD, were in the range of 0.41-1.24% (n = 3) and 2.17-3.17% (n = 9), respectively. The LOD was 0.19 ng, and the LOQ was 0.57 ng. Accuracy, calculated as percentage recovery, was between 98.98 and 101.96%, with a RSD not higher than 1.52%. The method was selective for the active principle tested. In conclusion, the method is useful for quantitative determination of carbamazepine in human serum. © 2009 Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim.
Más información
Título según WOS: | Instrumental planar chromatographic method for determination of carbamazepine in human serum |
Título según SCOPUS: | Instrumental planar chromatographic method for determination of carbamazepine in human serum |
Título de la Revista: | JOURNAL OF SEPARATION SCIENCE |
Volumen: | 32 |
Número: | 9 |
Editorial: | WILEY-V C H VERLAG GMBH |
Fecha de publicación: | 2009 |
Página de inicio: | 1454 |
Página final: | 1458 |
Idioma: | English |
URL: | http://doi.wiley.com/10.1002/jssc.200800675 |
DOI: |
10.1002/jssc.200800675 |
Notas: | ISI, SCOPUS |